System suitability parameters | Â |
 Retention time (min) | 2.9 |
 Asymmetric factor (T) | 0.8 |
 Selectivity (α) | 1.2 |
 Resolution (Rs) | 4.3 |
 Number of theoretical plates (N) | 20,927 |
Validation parameters | |
 Accuracy (mean ± SD)a | 100.7 ± 1.0 |
 Precision (%RSD) | |
  Repeatability b | 0.5 |
  Intermediate precision c | 1.6 |
Linearity | |
 Regression equation | Y = 1.79 + 6.81 |
 Correlation coefficient (r) | 0.9997 |
 Range (µg mL−1) | 1–50 |
 LOD (µg mL−1) d | 0.2 |
 LOQ (µg mL−1) e | 0.3 |
 Robustness (mean ± SD) f | 0.8 ± 0.01 |